On Biocompatibility and Stability of Transversal Intrafascicular Multichannel Electrodes—TIME

Thomas Stieglitz, Tim Boretius, Paul Čvančara, David Guiraud, Thomas Guiho, Victor Manuel Lopez-Alvarez, Xavier Navarro

Producción científica: Capítulo del libroCapítuloInvestigaciónrevisión exhaustiva

4 Citas (Scopus)

Resumen

© Springer International Publishing AG 2017. Transversal intrafascicular multichannel electrodes (TIME) have been developed to interface with peripheral nerves after upper limb amputation. Intended use is the electrical stimulation of the median and ulnar nerve to deliver sensory feedback during phantom limb pain treatment and artificial hand control. Miniaturized electrode arrays were developed on polyimide substrates with thin film metallization using sputtered iridium oxide as electrode coating. Here, we report on the essential requirements including biocompatibility, mechanical and stimulation stability that have been investigated before permission was granted by the legal authorities to conduct subchronic first-in-man clinical trials. Explants have been investigated to identify possible first failure points and optimize the devices for chronic implantation.
Idioma originalInglés
Título de la publicación alojadaBiosystems and Biorobotics
Páginas731-735
Número de páginas4
Volumen15
ISBN (versión digital)2195-3570
DOI
EstadoPublicada - 1 ene 2017

Huella

Profundice en los temas de investigación de 'On Biocompatibility and Stability of Transversal Intrafascicular Multichannel Electrodes—TIME'. En conjunto forman una huella única.

Citar esto