An electrical readout system of microarrays comprising an array of conductimetric transducers, which enabled multiplexed detection of up to 36 biological events on the same substrate, is reported in this thesis. Similarly to fluorescent readout counterparts, regular glass slides were applied for carrying out the microarray. However, unlike them, the presented system is compact and requires a simple and inexpensive instrumentation, thus being ideally suited for deployment of bioarray detection approaches. Other compact electrical readout systems previously reported are based on the immobilization of the specific bioreceptors on the transducer surface, which make them single-use and thus expensive. Our system combines the merits of both previous electrical systems and the fluorescence scanner approaches._x000D_ An array of 40 μm-pitch gold interdigitated electrode pairs (IDEs) was fabricated on glass substrates. Conductimetric measurements were carried out with these transducers by applying urease-based (urease labeled) immunoassay reactions. The hydrolysis of urea catalyzed by urease produces ionic species in solution that increase its conductivity. In order to carry out such detection scheme, the bioarray was developed on a regular glass slide substrate, which was then placed over the IDE microarray leaving a fixed gap and aligned so that each spot of the bioarray faced one IDE._x000D_ Therein, the droplets of urea solution were placed in the gap to contact the IDEs and the corresponding spots of the bioarray. However, in the development of the first approach, some major drawbacks related to droplet evaporation and chemical cross-talk between adjacent droplets made the system to be far from achieving the performance of fluorescent detection approaches. Moreover, a microdispenser was required to accurately position the droplets._x000D_ In a second approach, we found the solution for circumventing these drawbacks and the demonstration of the potential of the system as a real competitor against fluorescence based systems in terms of analytical performance. An array structure of 36 PDMS microwells having a volume of 1.2 μL were fabricated by molding and positioned over the IDE array. The geometry of each microwell includes an O-ring like structure and an empty area around it, which enabled first, to easily fill it without using any specific instrumentation and second, to ensure its liquid tightness once the bioarray was positioned over it. This set-up was demonstrated to completely avoid the evaporation and chemical cross-talk issues._x000D_ Finally, a third approach is presented, where the readout automatized by means of a microfluidic structure that includes the PDMS microwells together with those fluidic components that enabled the automatic filling with the measuring solution as well as the cleaning step of the system once a microarray was measured. Working in this way improves the throughput of the electrical readout system as well as enhanced its reliability by minimizing the manipulation by the user.
| Date of Award | 19 Dec 2013 |
|---|
| Original language | Spanish |
|---|
| Supervisor | César Fernández Sánchez (Director) & Antoni Baldi Coll (Director) |
|---|
Sistema de Lectura eléctrica de Microarrays Protéomicos.
Bonilla Aguilar, D. (Author). 19 Dec 2013
Student thesis: Doctoral thesis
Bonilla Aguilar, D. (Author), Fernández Sánchez, C. (Director) & Baldi Coll, A. (Director),
19 Dec 2013Student thesis: Doctoral thesis
Student thesis: Doctoral thesis