Estudio biomecánico de las placas dinámicas cervicales mediante modelos de elementos finitos y ensayos biomecánicos

Student thesis: Doctoral thesis

Abstract

Screwed plates are used in cervical surgeries to increase the fusion rate of the bone graft placed between two vertebrae, which design has evolved to overcome their disadvantages. In the initial models, the screws were not firmly attached to the plates, with risk of loosening of the same; this was addressed by firmly attaching the screws to the plates, but unfortunately this modifies the load distribution through the bone graft (following Wolff’s Law, compression is useful to foster bone formation). This is the reason why dynamic plates evolved, which allow graft load at the same time that guarantees enough stability for the fusion. Dynamicity can be achieved by rotation (toggling) of the screws, or by traslation of the screws or the plate itself; this can be so in one or two directions. On the other hand, during the fusion process there is a possibility of graft shortening; and the static plates may not be able to adapt to this situation. To the best of our knowledge, there is no study comparing the biomechanical behaviour of a bidirectional dynamic plate with its static counterpart, in a in vitro corpectomy model (the plate may not be necessary with discectomy). The purpose of this study is to investigate whether the supposed superiority of the dynamic plate can be certainly attributable to the intrinsic dynamism of the plate itself, or could be due to the plate design. For this purpose, both biomechanical essays and finite element models will be used; the latter are computer designs which combine geometric and physic elements. After the simulations with the plate in either static or dynamic work configuration, and with full graft length or simulated shortening, the results of the study show that the static plates bear more stress than the dynamic ones (which could theoretically in the long run cause implant loosening or breakage); and this increases as time goes by and the graft shortens, so the static plates are unable to effectively transfer the load. On the contrary, dynamic plates can transfer higher loads to the graft, and are also capable to adapt to the modified geometry after graft shortening, and allow the continuation of load transmission to the graft. For this reason, in those cases when plates are needed, it would be advisable to use traslational dynamic plates, as they have demonstrated in vitro the higher load transmission to the graft as well as less tension to the plate, which could eventually cause hardware complications.
Date of Award16 Jan 2023
Original languageSpanish
SupervisorFerran Pellise Urquiza (Director)

Cite this

'