Tendon transfer fixation in the foot and ankle: A biomechanical study

Susana Núñez-Pereira, Daniel Pacha-Vicente, Manuel Llusá-Pérez, Joan Nardi-Vilardaga

Research output: Contribution to journalArticleResearchpeer-review

14 Citations (Scopus)

Abstract

Background: Tendon transfers are often used in foot and ankle surgery. Different fixation devices and techniques have been described. The most recently developed ones are bone anchors and interference screws. Materials and Methods: A biomechanical study was designed to compare tendon transfer fixation, using Corkscrew bone anchors 5 x 15.5 mm and Biointerference screws (8 x 23 mm). Fifteen fresh cadaver specimens underwent both fixation techniques for split anterior tibial tendon transfer at the cuboid bone. All the specimens underwent standardized X-rays in order to evaluate mineral bone density using a standardized measurement system. All were tested until maximal load to failure. Results: The ultimate load to failure of the tendon secured to the cuboid using anchors was 103 N (SD, 52), compared with 150 N (SD, 68) for tendons secured to the bone with interference screws (p = 0.003). No difference was found between the techniques that could be related to bone density. Conclusion: Interference screws provided greater strength than bone anchors. Clinical Relevance: This study demonstrated increased strength in securing bone to tendon in vitro for SPLATT tendon transfer with interference screws as compared to suture anchors. © 2009 by the American Orthopaedic Foot & Ankle Society.
Original languageEnglish
Pages (from-to)1207-1211
JournalFoot and Ankle International
Volume30
Issue number12
DOIs
Publication statusPublished - 1 Dec 2009

Keywords

  • Bone anchors
  • Interference screws
  • Tendon transfer fixation

Fingerprint Dive into the research topics of 'Tendon transfer fixation in the foot and ankle: A biomechanical study'. Together they form a unique fingerprint.

Cite this