Simulation and experimental verification of ambient neutron doses in a pencil beam scanning proton therapy room as a function of treatment plan parameters

Olivier Van Hoey*, Liliana Stolarczyk, Jan Lillhök, Linda Eliasson, Natalia Mojzeszek, Malgorzata Liszka, Ali Alkhiat, Vladimir Mares, François Trompier, Sebastian Trinkl, Immaculada Martínez-Rovira, Maite Romero-Expósito, Carles Domingo, Ondrej Ploc, Roger Harrison, Pawel Olko

*Corresponding author for this work

Research output: Contribution to journalArticleResearchpeer-review

1 Citation (Scopus)

Abstract

Out-of-field patient doses in proton therapy are dominated by neutrons. Currently, they are not taken into account by treatment planning systems. There is an increasing need to include out-of-field doses in the dose calculation, especially when treating children, pregnant patients, and patients with implants. In response to this demand, this work presents the first steps towards a tool for the prediction of out-of-field neutron doses in pencil beam scanning proton therapy facilities. As a first step, a general Monte Carlo radiation transport model for simulation of out-of-field neutron doses was set up and successfully verified by comparison of simulated and measured ambient neutron dose equivalent and neutron fluence energy spectra around a solid water phantom irradiated with a variation of different treatment plan parameters. Simulations with the verified model enabled a detailed study of the variation of the neutron ambient dose equivalent with field size, range, modulation width, use of a range shifter, and position inside the treatment room. For future work, it is planned to use this verified model to simulate out-of-field neutron doses inside the phantom and to verify the simulation results by comparison with previous in-phantom measurement campaigns. Eventually, these verified simulations will be used to build a library and a corresponding tool to allow assessment of out-of-field neutron doses at pencil beam scanning proton therapy facilities.

Original languageEnglish
Article number903537
Number of pages22
JournalFrontiers in Oncology
Volume12
DOIs
Publication statusPublished - 8 Sep 2022

Keywords

  • Monte Carlo simulations
  • Neutron doses
  • Neutron measurements
  • Out-of-field neutron doses in radiation therapy
  • Pencil beam scanned proton therapy
  • Proton therapy

Fingerprint

Dive into the research topics of 'Simulation and experimental verification of ambient neutron doses in a pencil beam scanning proton therapy room as a function of treatment plan parameters'. Together they form a unique fingerprint.

Cite this