Sequential flow-injection determinations of calcium and magnesium in waters

J. Alonso, J. Bartroli, J. L.F.C. Lima, A. A.S.C. Machado

Research output: Contribution to journalArticleResearchpeer-review

44 Citations (Scopus)

Abstract

A tubular PVC membrane electrode for calcium without inner reference solution and a device for location of the reference electrode are described. In the flow-injection system, calcium is determined potentiometrically and then magnesium is determined by atomic absorption spectrometry. The electrode provides linear response to calcium in the range 5 × 10-5/2-10-1 M. On-line dilution of the sample allows magnesium determination in the range 0/2-10 mg l-. Flow rates between 3 and 6 ml min-1 are possible. The sampling frequency is 60/2-90 h-1. © 1986.
Original languageEnglish
Pages (from-to)503-508
JournalAnalytica Chimica Acta
Volume179
Issue numberC
DOIs
Publication statusPublished - 1 Jan 1986

Fingerprint Dive into the research topics of 'Sequential flow-injection determinations of calcium and magnesium in waters'. Together they form a unique fingerprint.

Cite this