Nitrate ion-selective electrode as reference electrode for flow-injection analysis

Research output: Contribution to journalArticleResearchpeer-review

8 Citations (Scopus)

Abstract

The development of a flow-injection system for pH determination using a pH ion-sensitive field-effect transistor in conjuction with a tubular flor-through nitrate ion-selective electrode (ISE) as a reference electrode is described. With this system, the linear working range extends from pH 1 to 13 with a sensitivity of 56 mV per pH unit and a relative standard deviation of 0.4%. A stable baseline and good respose are obtained by placing the tubular ISE in a closed branch of the main channel. © 1992.
Original languageEnglish
Pages (from-to)419-423
JournalAnal. Chim. Acta
Volume261
Issue number1-2
DOIs
Publication statusPublished - 25 May 1992

Keywords

  • Flow system
  • Ion-selective electrodes
  • Ion-sensitive field -effect transistors
  • Nitrate
  • pH
  • Potentiometry