TY - JOUR
T1 - Multi-center study of inter-rater reproducibility, image quality, and diagnostic accuracy of CZT versus conventional SPECT myocardial perfusion imaging
AU - Mansour, Nabeel
AU - Nekolla, Stephan G.
AU - Reyes, Eliana
AU - Angelidis, George
AU - Georgoulias, Panagiotis
AU - Anagnostopoulos, Constantinos
AU - Bravo, Paco
AU - Bruno, Isabella
AU - Flotats, Albert
AU - Fuentes-Ocampo, Francisco
AU - Sciagrà, Roberto
AU - Keng, Felix
AU - Kessler, Lukas
AU - Papathanasiou, Maria
AU - Soman, Prem
AU - Rischpler, Christoph
PY - 2022
Y1 - 2022
N2 - Cadmium-zinc-telluride (CZT)-based detectors exhibit higher diagnostic sensitivity in myocardial perfusion imaging (MPI) than conventional Anger-MPI for detection of coronary artery disease (CAD); however, reduced specificity and diagnostic accuracy of CZT-MPI were observed. This study aims to compare these different camera systems and to examine the degree of inter-rater reproducibility among readers with varying experience in MPI. 83 patients who underwent double stress/rest examinations using both a CZT and conventional SPECT cameras within one visit were included. Anonymized and randomized MPI-images were distributed to 15 international readers using a standardized questionnaire. Subsequent coronary angiography findings of ten patients served as a reference for analysis of sensitivity and specificity. Image quality was significantly better in CZT-MPI with significantly lower breast attenuation (P < 0.05). CZT-MPI exhibited higher sensitivity than Anger-MPI (87.5% vs. 62.5%) and significantly reduced specificity (40% vs. 100%). Readers experienced with both camera systems had the highest inter-rater agreement indicating higher reproducibility (CZT 0.54 vs. conv. 0.49, P < 0.05). Higher diagnostic sensitivity of CZT-MPI offers advantages in detection of CAD yet potentially of at the cost of reduced specificity, therefore it requires special training and a differentiated evaluation approach, especially for non-experienced readers with such camera systems. The online version contains supplementary material available at 10.1007/s12350-022-03054-w.
AB - Cadmium-zinc-telluride (CZT)-based detectors exhibit higher diagnostic sensitivity in myocardial perfusion imaging (MPI) than conventional Anger-MPI for detection of coronary artery disease (CAD); however, reduced specificity and diagnostic accuracy of CZT-MPI were observed. This study aims to compare these different camera systems and to examine the degree of inter-rater reproducibility among readers with varying experience in MPI. 83 patients who underwent double stress/rest examinations using both a CZT and conventional SPECT cameras within one visit were included. Anonymized and randomized MPI-images were distributed to 15 international readers using a standardized questionnaire. Subsequent coronary angiography findings of ten patients served as a reference for analysis of sensitivity and specificity. Image quality was significantly better in CZT-MPI with significantly lower breast attenuation (P < 0.05). CZT-MPI exhibited higher sensitivity than Anger-MPI (87.5% vs. 62.5%) and significantly reduced specificity (40% vs. 100%). Readers experienced with both camera systems had the highest inter-rater agreement indicating higher reproducibility (CZT 0.54 vs. conv. 0.49, P < 0.05). Higher diagnostic sensitivity of CZT-MPI offers advantages in detection of CAD yet potentially of at the cost of reduced specificity, therefore it requires special training and a differentiated evaluation approach, especially for non-experienced readers with such camera systems. The online version contains supplementary material available at 10.1007/s12350-022-03054-w.
KW - Cadmium-zinc-telluride
KW - Coronary artery disease
KW - Myocardial perfusion imaging
KW - Single-photon emission tomography
U2 - 10.1007/s12350-022-03054-w
DO - 10.1007/s12350-022-03054-w
M3 - Article
C2 - 35799039
SN - 1071-3581
VL - 30
SP - 528
EP - 539
JO - Journal of Nuclear Cardiology
JF - Journal of Nuclear Cardiology
ER -