Flow-through tubular PVC matrix membrane electrode without inner reference solution for flow injection analysis

S. Alegret, J. Alonso, J. Bartrolí, J. M. Paulís, J. L.F.C. Lima, A. A.S.C. Machado

Research output: Contribution to journalArticleResearchpeer-review

127 Citations (Scopus)

Abstract

A simply constructed tubular PVC matrix membrane electrode without inner reference solution suitable for flow injection analysis is described. The nitrate-selective electrode developed allows simple replacement of an exhausted membrane, or change of function with a new sensor membrane. The compact design ensures mechanical stability and the tubular configuration minimizes distortion of the hydrodynamic flow in the detector. Nitrate can be determined in the range 10-1-10-5 M, under different flow conditions, with a sampling frequency of about 120 h-1. © 1984.