Flow-through pH-ISFET detector for flow-injection analysis

S. Alegret, J. Alonso, J. Bartroli, J. Domenech, N. Jaffrezic-Renault, Y. Duvault-Herrera, Francesc Xavier Domenech Antunez

Research output: Contribution to journalArticleResearchpeer-review

19 Citations (Scopus)

Abstract

A simply constructed flow-through pH-ISFET detector suitable for flow-injection analysis is described. Its response characteristics are presented and discussed. The calibration plot is linear in the pH range 2-12; the slope is 53 mV per decade. © 1989.
Original languageEnglish
Pages (from-to)373-377
JournalAnalytica Chimica Acta
Volume222
Issue number1
DOIs
Publication statusPublished - 1 Jan 1989