A visual quality control scale for clinical arterial spin labeling images

S. M. Fallatah, F. B. Pizzini, B. Gomez-Anson, J. Magerkurth, E. De Vita, S. Bisdas, H. R. Jäger, H. J.M.M. Mutsaerts, X. Golay

    Research output: Contribution to journalArticleResearch

    Abstract

    © 2018, The Author(s). Background: Image-quality assessment is a fundamental step before clinical evaluation of magnetic resonance images. The aim of this study was to introduce a visual scoring system that provides a quality control standard for arterial spin labeling (ASL) and that can be applied to cerebral blood flow (CBF) maps, as well as to ancillary ASL images. Methods: The proposed image quality control (QC) system had two components: (1) contrast-based QC (cQC), describing the visual contrast between anatomical structures; and (2) artifact-based QC (aQC), evaluating image quality of the CBF map for the presence of common types of artifacts. Three raters evaluated cQC and aQC for 158 quantitative signal targeting with alternating radiofrequency labelling of arterial regions (QUASAR) ASL scans (CBF, T1 relaxation rate, arterial blood volume, and arterial transient time). Spearman correlation coefficient (r), intraclass correlation coefficients (ICC), and receiver operating characteristic analysis were used. Results: Intra/inter-rater agreement ranged from moderate to excellent; inter-rater ICC was 0.72 for cQC, 0.60 for aQC, and 0.74 for the combined QC (cQC + aQC). Intra-rater ICC was 0.90 for cQC; 0.80 for aQC, and 0.90 for the combined QC. Strong correlations were found between aQC and CBF maps quality (r = 0.75), and between aQC and cQC (r = 0.70). A QC score of 18 was optimal to discriminate between high and low quality clinical scans. Conclusions: The proposed QC system provided high reproducibility and a reliable threshold for discarding low quality scans. Future research should compare this visual QC system with an automatic QC system.
    Original languageEnglish
    Article number45
    JournalEuropean radiology experimental
    Volume2
    DOIs
    Publication statusPublished - 1 Dec 2018

    Keywords

    • Arterial spin labelling
    • Magnetic resonance imaging
    • Perfusion imaging
    • Quality control

    Fingerprint Dive into the research topics of 'A visual quality control scale for clinical arterial spin labeling images'. Together they form a unique fingerprint.

  • Cite this

    Fallatah, S. M., Pizzini, F. B., Gomez-Anson, B., Magerkurth, J., De Vita, E., Bisdas, S., Jäger, H. R., Mutsaerts, H. J. M. M., & Golay, X. (2018). A visual quality control scale for clinical arterial spin labeling images. European radiology experimental, 2, [45]. https://doi.org/10.1186/s41747-018-0073-2