T1 - Waves of Precision :
T2 - A Practical Guide for Reviewing New Tools to Evaluate Mechanical In-Exsufflation Efficacy in Neuromuscular Disorders
AU - Chatwin, Michelle
AU - Sancho, Jesus
AU - Luján, Manel
AU - Andersen, Tiina
AU - Winck, Joao-Carlos
PY - 2024
Y1 - 2024
N2 - Mechanical insufflation-exsufflation (MI-E) is essential for secretion clearance, especially in neuromuscular disorders. For the best outcomes, initiation of MI-E should be started at the correct time with regular evaluation to the response to treatment. Typically, cough peak flow has been used to evaluate cough effectiveness with and without MI-E. This review highlights the limitations of this and discussed other tools to evaluate MI-E efficacy in this rapidly developing field. Such tools include the interpretation of parameters (like pressure, flow and volumes) that derive from the MI-E device and external methods to evaluate upper airway closure. In this review we pinpoint the differences between different devices in the market and discuss new tools to better titrate MI-E and detect pathological responses of the upper airway. We discuss the importance of point of care ultrasound (POCUS), transnasal fiberoptic laryngoscopy and wave form analysis in this setting. To improve clinical practice newer generation MI-E devices should allow real-time evaluation of waveforms and standardize some of the derived parameters.
AB - Mechanical insufflation-exsufflation (MI-E) is essential for secretion clearance, especially in neuromuscular disorders. For the best outcomes, initiation of MI-E should be started at the correct time with regular evaluation to the response to treatment. Typically, cough peak flow has been used to evaluate cough effectiveness with and without MI-E. This review highlights the limitations of this and discussed other tools to evaluate MI-E efficacy in this rapidly developing field. Such tools include the interpretation of parameters (like pressure, flow and volumes) that derive from the MI-E device and external methods to evaluate upper airway closure. In this review we pinpoint the differences between different devices in the market and discuss new tools to better titrate MI-E and detect pathological responses of the upper airway. We discuss the importance of point of care ultrasound (POCUS), transnasal fiberoptic laryngoscopy and wave form analysis in this setting. To improve clinical practice newer generation MI-E devices should allow real-time evaluation of waveforms and standardize some of the derived parameters.
KW - Mechanical in-exsufflation
KW - Peak cough flow
KW - Waveform analysis
KW - Upper airway obstruction
KW - Neuromuscular disorders