TY - JOUR
T1 - Towards a monolithically integrated microsystem based on the green tape ceramics technology for spectrophotometric measurements. Determination of chromium (VI) in water
AU - Alves-Segundo, Ricardo
AU - Ibañez-Garcia, Núria
AU - Baeza, Mireia
AU - Puyol, Mar
AU - Alonso-Chamarro, Julián
PY - 2011/1/1
Y1 - 2011/1/1
N2 - A general procedure is presented for the fabrication of miniaturized continuous flow analytical microsystems based on photometric detection using the low temperature co-fired ceramics technology. Optical elements such as light emitting diodes and photodiodes (all in the size of a few hundred micrometers) are integrated by means of an off-chip approach. A simple procedure is demonstrated to integrate a glass window after the ceramic sintering in order to minimize the decrease in sensitivity due to the reduction of the optical path length when scaling down. A flow cell with a shape of a bubble has been used to increase the area of the light beam. The device is robust, affordable, and small-sized. It enables absorbance measurements to be performed in-situ or for continuous monitoring of environmental samples. Specifically, a microsystem is introduced for colorimetric determination of chromium (VI) ion in waters based on the diphenylcarbazide reagent as a model. Under the optimized conditions, a linear response is obtained for the concentration range from 0.1 to 20 mg L-1, with a detection limit of 50 μg L-1. © 2010 Springer-Verlag.
AB - A general procedure is presented for the fabrication of miniaturized continuous flow analytical microsystems based on photometric detection using the low temperature co-fired ceramics technology. Optical elements such as light emitting diodes and photodiodes (all in the size of a few hundred micrometers) are integrated by means of an off-chip approach. A simple procedure is demonstrated to integrate a glass window after the ceramic sintering in order to minimize the decrease in sensitivity due to the reduction of the optical path length when scaling down. A flow cell with a shape of a bubble has been used to increase the area of the light beam. The device is robust, affordable, and small-sized. It enables absorbance measurements to be performed in-situ or for continuous monitoring of environmental samples. Specifically, a microsystem is introduced for colorimetric determination of chromium (VI) ion in waters based on the diphenylcarbazide reagent as a model. Under the optimized conditions, a linear response is obtained for the concentration range from 0.1 to 20 mg L-1, with a detection limit of 50 μg L-1. © 2010 Springer-Verlag.
KW - Chromium (VI)
KW - Diphenylcarbazide
KW - Low-temperature co-fired ceramics
KW - Microanalyzer
KW - Photometric detection
U2 - 10.1007/s00604-010-0459-1
DO - 10.1007/s00604-010-0459-1
M3 - Article
SN - 0026-3672
VL - 172
SP - 225
EP - 232
JO - Microchimica Acta
JF - Microchimica Acta
IS - 1
ER -