TY - JOUR
T1 - Testing of a novel automated point-of-care analyzer for blood ammonium monitoring in a clinical setting
AU - Rebollo Calderón, Beatriz
AU - Calvo-Lopez, Antonio
AU - Ormazábal, Aida
AU - Artuch, R
AU - Rosell Ferrer, Javier
AU - Alonso Chamarro, Julián
AU - Puyol Bosch, M. del Mar
N1 - Publisher Copyright:
© The Author(s) 2025.
PY - 2025/6
Y1 - 2025/6
N2 - Certain diseases are marked by elevated ammonium levels in the blood, a condition known as hyperammonemia. Prompt detection and medical intervention are crucial to prevent potentially fatal outcomes. Therefore, ammonium levels should be monitored regularly, typically in referral hospitals where specialized and costly equipment is available. Although compact commercial devices are available for this purpose, none of them meet all the technical and analytical requirements needed for direct blood analysis, and current reported strategies have not been validated with enough samples to confirm results reliably. We present a robust and reliable automated point-of-care (POC) analyzer for the potentiometric determination of ammonium in blood. Comprising three computer-controlled modules-fluid management, detection, and data acquisition and transmission-this system combines portability, ease of use, and affordability. It can directly measure untreated blood samples, significantly reducing analysis time. Fully automated, it operates unsupervised with minimal lab personnel intervention. Analytical quality parameters include 5% RSD repeatability (n = 8), a limit of detection of 24 μM, a working range of 30-1000 µM and a sample volume of 215 µL. Successfully implemented in a hospital for 2 months, it analyzed 238 blood samples in parallel with the hospital's reference method showing comparable results (paired t-test, Passing-Bablok regression and Bland-Altman Plot) and randomly distributed errors, with a 4% accuracy calculated as mean error. Results indicate the POC analyzer effectiveness and reliability in a clinical setting compared to currently reported or commercially available equipment, being suitable for bedside monitoring of conditions associated with hyperammonemia in healthcare centers, including emergency rooms and clinics in developing countries.
AB - Certain diseases are marked by elevated ammonium levels in the blood, a condition known as hyperammonemia. Prompt detection and medical intervention are crucial to prevent potentially fatal outcomes. Therefore, ammonium levels should be monitored regularly, typically in referral hospitals where specialized and costly equipment is available. Although compact commercial devices are available for this purpose, none of them meet all the technical and analytical requirements needed for direct blood analysis, and current reported strategies have not been validated with enough samples to confirm results reliably. We present a robust and reliable automated point-of-care (POC) analyzer for the potentiometric determination of ammonium in blood. Comprising three computer-controlled modules-fluid management, detection, and data acquisition and transmission-this system combines portability, ease of use, and affordability. It can directly measure untreated blood samples, significantly reducing analysis time. Fully automated, it operates unsupervised with minimal lab personnel intervention. Analytical quality parameters include 5% RSD repeatability (n = 8), a limit of detection of 24 μM, a working range of 30-1000 µM and a sample volume of 215 µL. Successfully implemented in a hospital for 2 months, it analyzed 238 blood samples in parallel with the hospital's reference method showing comparable results (paired t-test, Passing-Bablok regression and Bland-Altman Plot) and randomly distributed errors, with a 4% accuracy calculated as mean error. Results indicate the POC analyzer effectiveness and reliability in a clinical setting compared to currently reported or commercially available equipment, being suitable for bedside monitoring of conditions associated with hyperammonemia in healthcare centers, including emergency rooms and clinics in developing countries.
KW - Potentiometry
KW - Hyperammonemia
KW - Inborn error of metabolism
KW - Point-of-care
KW - Ammonium selective electrodes
KW - Automation
KW - Limit of Detection
KW - Reproducibility of Results
KW - Ammonium Compounds/blood
KW - Humans
KW - Equipment Design
KW - Potentiometry/instrumentation
KW - Point-of-Care Systems
UR - http://www.scopus.com/inward/record.url?scp=105003123596&partnerID=8YFLogxK
UR - https://www.mendeley.com/catalogue/e63a5f9a-b61f-3793-a10b-1a9db9add011/
U2 - 10.1007/s00216-025-05879-z
DO - 10.1007/s00216-025-05879-z
M3 - Article
C2 - 40266320
SN - 1618-2642
VL - 417
SP - 3477
EP - 3485
JO - Analytical and Bioanalytical Chemistry
JF - Analytical and Bioanalytical Chemistry
M1 - E31796
ER -