T1 - Strategies for constructing the calibration set for a near infrared spectroscopic quantitation method
AU - Blanco, M.
AU - Romero, M. A.
AU - Alcalà, M.
PY - 2004/10/20
Y1 - 2004/10/20
N2 - Three strategies for the construction of calibration sets have been tried, with the objective to develop and to validate a NIR quantitation method. The first two approaches consist of the use of two types of samples, named: samples of laboratory obtained by mixing the ingredients that compose the drug, and doped samples obtained by under- and over-dosed production samples. In order to improve the prediction results, production samples have been added to each calibration model. The ensuing models were validated with a view to determine their fitness for purpose. However, spectral differences between the laboratory samples and doped samples resulted in spurious predictions in quantifying samples of one type using the model developed from samples of the other. Such differences were studied in depth and a third procedure has been proposed, based on a calibration model constructed with an unique type of sample (laboratory sample) for later to correct it with a few doped samples. This corrected model has a good predictive ability on production samples. © 2004 Elsevier B.V. All rights reserved.
AB - Three strategies for the construction of calibration sets have been tried, with the objective to develop and to validate a NIR quantitation method. The first two approaches consist of the use of two types of samples, named: samples of laboratory obtained by mixing the ingredients that compose the drug, and doped samples obtained by under- and over-dosed production samples. In order to improve the prediction results, production samples have been added to each calibration model. The ensuing models were validated with a view to determine their fitness for purpose. However, spectral differences between the laboratory samples and doped samples resulted in spurious predictions in quantifying samples of one type using the model developed from samples of the other. Such differences were studied in depth and a third procedure has been proposed, based on a calibration model constructed with an unique type of sample (laboratory sample) for later to correct it with a few doped samples. This corrected model has a good predictive ability on production samples. © 2004 Elsevier B.V. All rights reserved.
KW - Calibration
KW - Near infrared spectroscopy
KW - NIR
KW - Pharmaceutical analysis
UR - https://www.scopus.com/pages/publications/4544354367
U2 - 10.1016/j.talanta.2004.03.027
DO - 10.1016/j.talanta.2004.03.027
M3 - Article
SN - 0039-9140
VL - 64
SP - 597
EP - 602
JO - Talanta
JF - Talanta
IS - 3
ER -