Pesticide determination in tap water and juice samples using disposable amperometric biosensors made using thick-film technology

Miquel Albareda-Sirvent, Arben Merkoçi, Salvador Alegret

Producció científica: Contribució a revistaArticleRecercaAvaluat per experts

86 Cites (Scopus)

Resum

A new design of biosensor strips for pesticide analysis is presented. The developed biosensors integrate a photolithographic conducting copper track, graphite-epoxy composite applied by screen-printing and enzyme (AChE or BChE) immobilised manually by crosslinking with glutaraldehyde. Detection limits in the order of 10-10 to 10-11 M pesticides (paraoxon and carbofuran) have been achieved in standard solutions. Pesticide analysis have been realised on spiked real samples of tap water and fruit juice, with recovering percentages close to 100%, corroborating that the biosensor is sensitive enough to detect pesticides in these matrices. © 2001 Elsevier Science B.V. All rights reserved.
Idioma originalAnglès
Pàgines (de-a)35-44
RevistaAnalytica Chimica Acta
Volum442
Número1
DOIs
Estat de la publicacióPublicada - 31 d’ag. 2001

Fingerprint

Navegar pels temes de recerca de 'Pesticide determination in tap water and juice samples using disposable amperometric biosensors made using thick-film technology'. Junts formen un fingerprint únic.

Com citar-ho