TY - JOUR
T1 - Monte Carlo simulations of a microstructured silicon detector with high efficiency for thermal neutrons
AU - Guardiola, C.
AU - Fleta, C.
AU - Amgarou, K.
AU - Lozano, M.
PY - 2012/6/1
Y1 - 2012/6/1
N2 - In this work we present a design optimization of a high efficiency neutron detector with MCNPX simulations. The detector is based on a silicon structure incorporating a boron-based converter to detect thermal neutrons by means of the 10B(n,α)7Li nuclear reaction. In order to improve the efficiency of planar detectors, limited to 5%, we have designed a perforated detector based on microstructures etched inside the silicon bulk, which would considerably increase this value. Each one of these microstructures is a micro-channel filled with the converter that provides a high surface-volume contact ratio between the converter and the sensitive silicon bulk. The main parameters optimized with the simulations are micro-channel width and depth, width of the silicon wall between micro-channels, and converter density, all comprised within technologically viable ranges. The results show that the detector could achieve up to 50% thermal neutron efficiency for a realistic prototype that can be produced with MEMS fabrication techniques. © 2012 IOP Publishing Ltd and Sissa Medialab srl.
AB - In this work we present a design optimization of a high efficiency neutron detector with MCNPX simulations. The detector is based on a silicon structure incorporating a boron-based converter to detect thermal neutrons by means of the 10B(n,α)7Li nuclear reaction. In order to improve the efficiency of planar detectors, limited to 5%, we have designed a perforated detector based on microstructures etched inside the silicon bulk, which would considerably increase this value. Each one of these microstructures is a micro-channel filled with the converter that provides a high surface-volume contact ratio between the converter and the sensitive silicon bulk. The main parameters optimized with the simulations are micro-channel width and depth, width of the silicon wall between micro-channels, and converter density, all comprised within technologically viable ranges. The results show that the detector could achieve up to 50% thermal neutron efficiency for a realistic prototype that can be produced with MEMS fabrication techniques. © 2012 IOP Publishing Ltd and Sissa Medialab srl.
KW - Detector modeling and simulations I (interaction of radiation with matter, interaction of photons with matter, interaction of hadrons with matter, etc)
KW - Neutron detectors (cold, thermal, fast neutrons)
KW - Solid state detectors
UR - https://www.scopus.com/pages/publications/84862905789
U2 - 10.1088/1748-0221/7/06/T06003
DO - 10.1088/1748-0221/7/06/T06003
M3 - Article
SN - 1748-0221
VL - 7
JO - Journal of Instrumentation
JF - Journal of Instrumentation
IS - 6
M1 - T06003
ER -