TY - JOUR
T1 - Actual performance of mechanical ventilators in ICU: A multicentric quality control study
AU - Govoni, Leonardo
AU - Dellaca, Raffaele L.
AU - Peñuelas, Oscar
AU - Bellani, Giacomo
AU - Artigas, Antonio
AU - Ferrer, Miquel
AU - Navajas, Daniel
AU - Pedotti, Antonio
AU - Farré, Ramon
PY - 2012/12/19
Y1 - 2012/12/19
N2 - Even if the performance of a given ventilator has been evaluated in the laboratory under very well controlled conditions, inappropriate maintenance and lack of long-term stability and accuracy of the ventilator sensors may lead to ventilation errors in actual clinical practice. The aim of this study was to evaluate the actual performances of ventilators during clinical routines. A resistance (7.69 cmH2O/L/s) - elastance (100 mL/cmH2O) test lung equipped with pressure, flow, and oxygen concentration sensors was connected to the Y-piece of all the mechanical ventilators available for patients in four intensive care units (ICUs; n = 66). Ventilators were set to volume-controlled ventilation with tidal volume = 600 mL, respiratory rate = 20 breaths/minute, positive end-expiratory pressure (PEEP) = 8 cmH2O, and oxygen fraction = 0.5. The signals from the sensors were recorded to compute the ventilation parameters. The average ± standard deviation and range (min-max) of the ventilatory parameters were the following: inspired tidal volume = 607 ± 36 (530-723) mL, expired tidal volume = 608 ± 36 (530-728) mL, peak pressure = 20.8 ± 2.3 (17.2-25.9) cmH2O, respiratory rate = 20.09 ± 0.35 (19.5-21.6) breaths/minute, PEEP = 8.43 ± 0.57 (7.26-10.8) cmH2O, oxygen fraction = 0.49 ± 0.014 (0.41-0.53). The more error-prone parameters were the ones related to the measure of flow. In several cases, the actual delivered mechanical ventilation was considerably different from the set one, suggesting the need for improving quality control procedures for these machines. © 2012 Govoni et al, publisher and licensee Dove Medical Press Ltd.
AB - Even if the performance of a given ventilator has been evaluated in the laboratory under very well controlled conditions, inappropriate maintenance and lack of long-term stability and accuracy of the ventilator sensors may lead to ventilation errors in actual clinical practice. The aim of this study was to evaluate the actual performances of ventilators during clinical routines. A resistance (7.69 cmH2O/L/s) - elastance (100 mL/cmH2O) test lung equipped with pressure, flow, and oxygen concentration sensors was connected to the Y-piece of all the mechanical ventilators available for patients in four intensive care units (ICUs; n = 66). Ventilators were set to volume-controlled ventilation with tidal volume = 600 mL, respiratory rate = 20 breaths/minute, positive end-expiratory pressure (PEEP) = 8 cmH2O, and oxygen fraction = 0.5. The signals from the sensors were recorded to compute the ventilation parameters. The average ± standard deviation and range (min-max) of the ventilatory parameters were the following: inspired tidal volume = 607 ± 36 (530-723) mL, expired tidal volume = 608 ± 36 (530-728) mL, peak pressure = 20.8 ± 2.3 (17.2-25.9) cmH2O, respiratory rate = 20.09 ± 0.35 (19.5-21.6) breaths/minute, PEEP = 8.43 ± 0.57 (7.26-10.8) cmH2O, oxygen fraction = 0.49 ± 0.014 (0.41-0.53). The more error-prone parameters were the ones related to the measure of flow. In several cases, the actual delivered mechanical ventilation was considerably different from the set one, suggesting the need for improving quality control procedures for these machines. © 2012 Govoni et al, publisher and licensee Dove Medical Press Ltd.
KW - Equipment and supplies
KW - Health care quality assessment
KW - Intravenous
KW - Medical devices
KW - Quality assurance
KW - Ventilation error
KW - Ventilator accuracy
UR - https://www.scopus.com/pages/publications/84871632483
U2 - 10.2147/MDER.S35864
DO - 10.2147/MDER.S35864
M3 - Article
SN - 1179-1470
VL - 5
SP - 111
EP - 119
JO - Medical Devices: Evidence and Research
JF - Medical Devices: Evidence and Research
IS - 1
ER -